Chapter 3: Testing Performance
To test single cursor voltage measurement accuracy with offset
44
Figure 3-3
Voltage Measurement Accuracy Equipment Setup
2 Press Default Setup to set the oscilloscope to default conditions.
3 Set all channels to DC using the Coupling key and to 1 M input impedance using the
Input key.
4 Using the mouse, enable the graphical interface.
5 Select Acquisition from the Setup menu.
6 Select Equivalent Time sampling mode. Enable Averaging. Set the # of Averages to 32,
using the numeric keypad dialog. Select Close.
See Figure 3-4.
Figure 3-4
Acquisition Setup for Voltage Accuracy Measurement
7 Use the following table for steps 8 through 15.
Select Equivalent Time
sampling mode
Enable averaging
Set the number of
averages to 32
Leave Memory Depth
on Automatic