© Agilent Technologies2000–2003 Agilent 3070 / 79000 DUT Power Supply Installation 72
Connecting the DUT Power Supply Cables
Programming the GPIB Address on Supplies with Front
Panels
From the front panel of the supply, program the GPIB
address by pressing
ADDR, entering the address, and
pressing
ENTER. Be certain not to duplicate a GPIB
address.
Table 9 shows the default GPIB addresses.
NOTE
If a supply is split between two modules, use the
device file address and GPIB address of the lower
numbered module.
Table 9 DUT power supply GPIB addresses
Power Supply Connection GPIB Address Device File Address
Module 0, ASRU channels 1-4 22 /dev/ps0
Module 1, ASRU channels 1-4 23 /dev/ps1
Module 2, ASRU channels 1-4 24 /dev/ps2
Module 3, ASRU channels 1-4 25 /dev/ps3
Module 0, ASRU channel 5 or 5-6 26 /dev/ps4
Module 0, ASRU channel 6 27 /dev/ps5
Module 1, ASRU channel 5 or 5-6 28 /dev/ps6
Module 1, ASRU channel 6 29 /dev/ps7
Module 2, ASRU channel 5 or 5-6 1 /dev/ps8
Module 2, ASRU channel 6 2 /dev/ps9
Module 3, ASRU channel 5 or 5-6 3 /dev/ps10
Module 3, ASRU channel 6 4 /dev/ps11