Agilent Technologies J-BERT N4903B Wartungshandbuch Seite 75

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 406
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 74
2
The following query returns the "instantaneous" data rate ratio (Actual DRR)
calculated from the counts obtained in the last deci-second. This value is
available even when accumulation is turned off.
[P]FETCh[:SENSe[1]]:DRRatio[:ALL][:FULL]:DELTa?
How to Evaluate Auto Re-Sync Count Results
You can use the following commands/queries to evaluate Auto Re-Sync Count
results:
1
The following query returns total number of Auto Re-Sync Count received in a
time interval.
[P]FETCh[:SENSe[1]]:ARSCount[:ALL][:FULL][:TOTal]?
2
The following query returns the “Instantaneous” number of auto re-sync count,
calculated from the counts obtained in the last deci- second. This value is
available even when accumulation is turned off.
[P]FETCh[:SENSe[1]]:ARSCount[:ALL][:FULL]:DELTa?
How to Evaluate Expected Filler Symbol Count Results
The following query returns total number of Expected Filler Symbol Count received
in a time interval.
[P]FETCh[:SENSe[1]]:EFSCount?
How to Evaluate Received Symbol Count Results
The following query returns total number of Received Symbol Count received in a
time interval.
[P]FETCh[:SENSe[1]]:RSCOunt?
Working with PCIe3 SKPOS
Working with PCIe 3.0 SKPOS – Concepts
The Serial BERT provides enhanced capability to perform a receiver (RX) jitter
tolerance test for PCIe3 (PCI
®
Express 3.0). In PCIe3, when RX detects incoming
data correctly, the pattern looped back has the "same" content except the length
Recommended Programming Techniques 3
Agilent J-BERT N4903B High-Performance Serial BERT 75
Seitenansicht 74
1 2 ... 70 71 72 73 74 75 76 77 78 79 80 ... 405 406

Kommentare zu diesen Handbüchern

Keine Kommentare