The query returns the present number.
MEAS:JTOL:CHAR(*):PARameter:SEARch[?]
IAgilentN490xJCharacterizationMeasurement.SearchAlgorithm
MEAS:JTOLerance:CHAR(*):PARameter:SEARch ULINear | ULOGarithmic |
DLINear | DLOGarithmic | BINary | EBINary | BINSPECIAL
MEAS:JTOL:CHAR(*):PAR:SEARch?
The command sets the search method to one of the following:
upwards linear (ULIN)
upwards logarithmic (ULOG)
downwards linear (DLIN)
downwards logarithmic (DLOG)
binary (BIN)
extended binary (EBIN)
The query returns the present setting.
MEAS:JTOL:CHAR(*):PAR:SEARch:ULINear:STEP[?]
IAgilentN490xJCharacterizationMeasurement.GetSearchParameter
MEAS:JTOLerance:CHAR(*):PARameter:SEARch:ULINear:STEP <NR3>
MEAS:JTOL:CHAR(*):PAR:SEARch:ULIN:STEP?
The command sets the jitter amplitude step size for an upwards linear search. The
unit is UI.
The query returns the present step size.
MEAS:JTOL:CHAR(*):PAR:SEARch:DLINear:STEP[?]
IAgilentN490xJCharacterizationMeasurement.SetSearchParameter/
GetSearchParameter
MEAS:JTOLerance:CHAR(*):PARameter:SEARch:DLINear:STEP <NR3>
5 SCPI Command Reference
358 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Seitenansicht 357