This command is kept for compatibility reasons. Due to the lack of relays, it is not
possible to disable the Serial BERT's output. This command has no effect. The
query will always return "1" = ON.
To force the Data Out and Clock Out ports to 0 V, use the command “OUTPut
[1]:CENTral[?] ” on page 135.
OUTPut[1]:TERMination[?]
IAgilentN490xPGOutVoltage.VTermination (IVI-compliant)
OUTPut[1]:TERMination <NR3>
OUTPut[1]:TERMination?
This command sets the data termination level of the pattern generator's Data Out
port. The response form returns the data termination level.
This command is only valid if the coupling is set to DC (see “OUTPut[1]:COUPling
[?] ” on page 136).
OUTPut[1]:WIDTh[?]
IAgilentBertPGPulse.Width
OUTPut[1]:WIDTh <NR3>
OUTPut[1]:WIDTh?
Sets the width or duration of a repetitive pulse (like in RZ or R1 signal modes). The
width value is given in seconds (<NR3>). Valid range is 25 ps (e.g. 25 e-12) to one
clock period minus 25 ps.
The query returns the current setting.
OUTPut[1]:EIDLe[:STATe]
IAgilent:EIDLe[:STATe]
OUTPut[1]:EIDLe[:STATe] 0 | 1 | OFF | ON
OUTPut[1]:EIDLe[:STATe]?
SCPI Command Reference 5
Agilent J-BERT N4903B High-Performance Serial BERT 139
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax