SOURce3:VOLTage:PROTection:RESet[?]
IAgilentN490xPGTrigVoltage.ProtectionReset (not IVI-compliant)
SOURce3:VOLTage:PROTection:RESet[?]
SOURce3:VOLTage:PROT:RES?
This command reset the protection circuit to normal operation at the Trigger Output
channel.
SOURce3:VOLTage:PROTection[:STATe][?]
IAgilentN490xPGTrigVoltage.LogicLevel (not IVI-compliant)
SOURce3:VOLTage:PROTection[:STATe] ON | OFF
SOURce3:VOLTage:PROT:STAT?
This command gets the state of the protection circuit at the Trigger Output channel.
The 'ON' state indicates a triggered protection circuit or in other words the output
is disabled and no signal comes out.
SOURce3:TRIGger:OUTPut:COUPling[?]
IAgilentN490xPGOutput.TerminationEnabled (IVI-compliant)
SOURce3:TRIGger:OUTPut:COUPling AC | DC
SOURce3:TRIGger:OUTPut:COUPling?
The command enables or disables the source of the termination voltage:
DC: Enables the termination voltage
AC: Disables the termination voltage
The query returns the current state.
SOURce3:TRIGger:OUTPut:TERMination[?]
IAgilentN490xPGOutVoltage.VTermination (IVI-compliant)
5 SCPI Command Reference
182 Agilent J-BERT N4903B High-Performance Serial BERT
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Syntax
Description
IVI-COM Equivalent
Seitenansicht 181
1 2 ... 177 178 179 180 181 182 183 184 185 186 187 ... 405 406