Agilent Technologies E6474A Betriebsanweisung Seite 114

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 224
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 113
114 Agilent E6474A User’s Guide
4 Use Your System
If your system is equipped with a navigator, perform the tests in
Table 5 on page 116 next. If you experience problems in setting
up the tests, refer to “Solving Problems with Your System" on
page 130.
Test 3 - Receiver Testing by Measuring the Noise Floor
Preparing the Hardware
1 Connect the system as shown in Figure 49.
2 Turn on the Agilent receiver.
NOTE
Use this procedure only if you do not have a signal generator and there is
no known channel in the system. Connect a 50-ohm load (customer
supplied) to the input of the receiver.
Figure 49 System Configuration for Measuring the Noise Floor
Seitenansicht 113
1 2 ... 109 110 111 112 113 114 115 116 117 118 119 ... 223 224

Kommentare zu diesen Handbüchern

Keine Kommentare