Agilent Technologies 1100 Series Bedienerhandbuch Seite 245

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 272
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 244
245
Control Module Screens for the Quaternary Pump
Screens available from the Records screen
Info Log Use the m-key to receive a pop-up menu, Select Info Log. A list of the last
events are listed. For troubleshooting reasons they can be printed or saved to
a file on the PCMCIA card (using the m-key for the context sensitive menu).
EMF (Early
Maintenance
Feedback)
Use the F1 key (EMF) to set EMF parameters. Choose menu item 1 (Setup
limits) to select the amount of pumped solvent or seal wear, when you want
to receive a warning for exceeded limits. Press F7 (Reset) to reset the
counters after exchanging parts that are subject to wear.
Seitenansicht 244
1 2 ... 240 241 242 243 244 245 246 247 248 249 250 ... 271 272

Kommentare zu diesen Handbüchern

Keine Kommentare