Repairing the Pump
Removing the Top Cover and Foam
Removing the Top Cover and Foam
Tools required Screwdriver Pozidriv #1
Preparations for this
procedure
Switch off quaternary pump at the main power switch.
Disconnect the solvent inlet and outlet tubes from the pump.
Beware of leaking solvents due to hydrostatic flow.
Remove the solvent cabinet from the quaternary pump.
Remove leak funnel with the waste tube from the top cover of the quaternary pump.
1 Remove the front cover by pressing the both
clip fasteners on both sides of the cover.
2 Unplug the power cord and move the lever
towards the power socket.
(1)
Clip fastener (1)
(2)
Lever