Making Measurements 4
N9340A User’s Guide 57
2 Set one source (signal generator) to 300 MHz
and the other source to 301 MHz. This will
define the frequency separation at 1 MHz. Set
both sources equal in amplitude, as measured by
the analyzer. In this example, they are both set
to –5 dBm.
3 Set the analyzer center frequency and span:
Press [PRESET]. (With Preset Type of Default)
Press [FREQ] > {Center Freq} > 300.5 > {MHz}.
Press [SPAN] > 5 > {MHz}.
4 Reduce the analyzer RBW until the distortion
products are visible:
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
5 Move the signal to the reference level:
Press [MARKER] > {Peak Search}.
Press [MARKER] > {Marker To} > {To Ref}.
6 Reduce the RBW until the distortion products
are visible:
Press [BW/SWP] > {RBW}, and reduce the RBW
using the knob, the arrow keys or the numeric
keypad.
7 Activate the second marker and place it on the
peak of the distortion product (beside the test
signal) using the Next Peak:
Press [MARKER] > {Delta}.
Press [MARKER] > {Peak Search} > {Next Left
(Right) PK}.
8 Measure the other distortion product:
Press [MARKER] > {Normal}.
NOTE
The coupler used should have a high isolation between
the two input ports to limit the source intermodulation.