Agilent Technologies N9340A Betriebsanweisung Seite 44

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 179
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 43
4System Setting
38 N9340A User’s Guide
Perform a Time Base Calibration
When the calibration function is triggered, the
current measurement is interrupted and a gauge
displays on the LCD. The gauge indicates
calibration action rather than calibration course, as
the calibration time is unpredictable. When the
calibration is finished, the LCD displays a
calibration, and the interrupted measurement
restarts.
Key Access: [SYS] > {More} > {Calibration}
NOTE
Time base calibration takes only a short time when the inner
temperature is stable. When the inner temperature is
increasing, calibration may take a long-time or will fail. If the
input reference signal is abnormal, the calibration cycle will
take a long and unpredictable time to exit, and the LCD will
display an error message.
Seitenansicht 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 178 179

Kommentare zu diesen Handbüchern

Keine Kommentare