6 SCPI Command Reference
140 N9340A User’s Guide
Turn Averaging On/Off
[:SENSe]:AVERage:TRACe 1|2|3|4|5|6[:STATe]
OFF|ON|0|1
[:SENSe]:AVERage:TRACe 1|2|3|4|5|6[:STATe]?
This command toggles averaging between off and
on. Averaging combines the value of successive
measurements to average out measurement
variations.
Set the Average Count
[:SENSe]:AVERage:TRACe1|2|3|4|5|6:COUNt <para>
[:SENSe]:AVERage:TRACe 1|2|3|4|5|6:COUNt?
This command specifies the number of
measurements that are combined.
*RST State: Off
Key Access: [TRACE] > {Average}
*RST Value: 100
Key Access: [TRACE] > {Average}