TRL/LRM Calibration
Overview
The TRL/LRM Calibration VBA Macro lets you perform the 2-, 3-, and 4-port TRL (Through-Reflect-Line),
LRL (Line-Reflect-Line), TRM (Through-Reflect-Match), and LRM (Line-Reflect-Match) calibrations for the
E5070B/E5071B ENA RF network analyzers. These calibration techniques are generally used in non-coaxial
environments such as in-fixture measurements and on-wafer measurements.
TRL calibration is performed by using the zero-length through, reflection (short, or open), and line standards.
When the non-zero-length through is used instead of the zero-length through standard, the calibration is called
LRL. TRM calibration is performed by using the zero-length through, reflection (short, or open), and match
(load) standards. When the non-zero-length through is used instead of the zero-length through standard, the
calibration is called LRM.
TRM and LRM calibrations are suitable for measurements at relatively low frequencies where the line
standards of the TRL and LRL calibrations can become physically too long for practical use. The TRL/LRM
Calibration VBA Macro allows you to perform calibration over a broad frequency range with a combination
of TRM (or LRM) and TRL (or LRL) calibrations while using up to three line standards.
Table 1. Key features (comparison with old version)
Function TRL Cal.
VBA Macro (old version)
TRL/LRM Cal.
VBA Macro
TRL and LRL calibrations
TRM and LRM calibrations
Multiport calibrations
Number of line standards
Setting negative delay values for cal. stds
Averaging for calibration measurements
Yes
No
Yes
Up to 2
No
No
Yes
Yes
Yes
Up to 3
Yes
Yes
NOTE
We recommend that you use the E5070B/E5071B firmware revision 3.5 and above.