Reflection measurement:
- Press Reflection.
- Connect the Reflection standard to test port 1, and press [ ]Reflection 1. The reflection
measurement is performed, and an asterisk appears.
- Connect the Reflection standard to test port 2, and press [ ]Reflection 2. The reflection
measurement is performed, and an asterisk appears.
Match measurement:
- Press Match.
- Connect the Match standard to test port 1 and press [ ]Match 1. The Match measurement is
performed, and an asterisk appears.
- Connect the Match standard to test port 2 and press [ ]Match 2. The Match measurement is
performed, and an asterisk appears.
Line1 measurement:
- Press Line1.
- Connect the Line1 standard between test ports 1 and 2, and press [ ]Line1 1-2. The Line1
measurement is performed, and an asterisk appears.
Line2 measurement:
- Press Line2.
- Connect the Line2 standard between the ports 1 and 2, and then press
[ ]Line2 1-2
. The Line2
measurement is performed, and an asterisk appears.
6-2) When all of the necessary calibration measurements are completed, the check marks are indicated on the
Thru/Reflection/Match/Lines keys, and the Update CalCoef key appears. Press the Update CalCoef key to
activate the measured calibration data. Confirm that the indication “Cor” appears on the display.
a) Press Update CalCoef to activate cal. data. b) “Cor is indicated on the display.
Figure 6. Update CalCoef key to activate calibration data
7. Finishing calibration
To leave TRL/LRM calibration, click on the close button [X] in the upper right hand corner of the main
control panel.