Agilent Technologies 8935 Series Bedienungsanleitung Seite 319

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 352
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 318
Chapter 1 319
GPIB Command Dictionary
SANalyzer subsystem
:MARKer[:NORMal]:NPEak
This command moves the marker to the next peak of the spectrum
analyzers measurement trace.
Syntax
SAN:MARK:NORM:NPE
Screen/field equivalent
SAN:MARK:NORM:NPE controls the Marker To, Next Peak field on
the Marker menu of the SPEC ANL screen when Norm/Delta field is set
to Norm on the Marker controls menu.
:MARKer[:NORMal]:NPLevel <real number>
:MARKer[:NORMal]:NPLevel?
These commands set/query the level of the next peak signal.
This command utilizes the “Real Number Setting Syntax” on page 299,
but does not use the :STATe command.
Syntax
SAN:MARK:NORM:NPL <real number>
SAN:MARK:NORM:NPL?
Screen/field equivalent
SAN:MARK:NORM:NPL controls the Level field on the SPEC ANL
screen when Norm/Delta field is set to Norm on the Marker controls
menu.
:MARKer[:NORMal]:PEAK
This command moves the marker to the highest peak of the signals on
the spectrum analyzer display trace.
Syntax
SAN:MARK:NORM:PEAK
Screen/field equivalent
SAN:MARK:NORM:PEAK controls the Marker To, Peak field on the
Marker menu of the SPEC ANL screen when Norm/Delta field is set to
Norm on the Marker controls menu.