Agilent Technologies 8935 Series Bedienungsanleitung Seite 141

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 352
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 140
Chapter 1 141
GPIB Command Dictionary
CDANalyzer subsystem
:IS2000:POWer:FAST:NAVG <integer>
:IS2000:POWer:FAST:NAVG?
IS-2000 only. These commands set/query the number of averages made
during fast power measurements.
Syntax
CDAN:IS2000:POW:FAST:NAVG <integer>
CDAN:IS2000:POW:FAST:NAVG?
Screen/field equivalent
CDAN:IS2000:POW:FAST:NAVG controls the Num Avgs field of the FP
Setup menu on the CODE DOMAIN screen when the CDMA standard
is set to “IS-2000”.
:IS2000:POWer:UNIT <string>
:IS2000:POWer:UNIT?
IS-2000 only. These commands set/query whether power measurements
are absolute power measurements or relative power measurements.
‘Abs’ - Absolute code domain power displays the power in each of the
128 Walsh channels, relative to the total power inside a 1.23 MHz
bandwidth centered at the tune frequency.
‘Rel’ - Relative code domain power displays the power in each of the
128 Walsh channels, relative to the power of the Pilot signal. (Pilot
power is approximately two-thirds of the total power.)
Syntax
CDAN:IS2000:POW:UNIT ‘Abs’
‘Rel’
CDAN:IS2000:POW:UNIT?
Screen/field equivalent
CDAN:IS2000:POW:UNIT controls the CD pwr unit field of the
Reference image on the CODE DOMAIN screen when the CDMA
standard is set to “IS-2000”.
Seitenansicht 140
1 2 ... 136 137 138 139 140 141 142 143 144 145 146 ... 351 352

Kommentare zu diesen Handbüchern

Keine Kommentare