32 Chapter 1
O:\Manuals\E6380A_Progguid\Book\Usehpib.fm
Using HP-IB
Getting Started
General HP-IB Programming Guidelines
The following guidelines should be considered when developing
programs which control the Test Set through HP-IB:
Guideline #1.
1. Bring the Test Set to its preset state using the front-panel
PRESET key. This initial step allows you to start developing
the measurement sequence with most fields in a known state.
2. Make the measurement manually using the front-panel
controls of the Test Set. Record, in sequential order, the
screens selected and the settings made within each screen. The
record of the screens selected and settings made in each screen
becomes the measurement procedure.
3. Record the measurement result(s).
4. Develop the program using the measurement procedure
generated in step 2. Be sure to start the programmatic
measurement sequence by bringing the Test Set to its preset
state using the *RST Common Command. As the
measurement procedure requires changing screens, use the
DISPlay command to select the desired screen followed by the
correct commands to set the desired field(s).
NOTE
When IBASIC programs are running the display is dedicated to the
IBASIC controller for program and graphics display. This means
instrument front panels are not displayed when an IBASIC program is
running. However, the DISPlay <screen> command causes all setting
and measurement fields in the <screen> to be accessible
programmatically. Attempting to read from a screen that has not been
made accessible by the DISPlay command will cause
HP-IB Error:
420 Query UNTERMINATED, or
HP-IB Error:
113 Undefined header
Make sure the desired measurement is in the ON state. This is
the preset state for most measurements. However, if a previous
program has set the state to OFF, the measurement will not be
available. Attempting to read from a measurement field that is
not in the ON state will cause HP-IB Error:
420 Query
UNTERMINATED.
5. If the trigger mode has been changed, trigger a reading.
NOTE
Triggering is set to FULL SETTling and REPetitive RETRiggering
after receipt of the *RST Common Command. These settings cause the
Test Set to trigger itself and a separate trigger command is not
necessary.