Chapter 8 233
IBASIC Controller
Interfacing to the IBASIC Controller using Serial Ports
Receive and Transmit Pacing
When receiving characters into the IBASIC Command Line field, the
Test Set’s microprocessor responds to each entry and no buffering is
required. Therefore, when using your PC or terminal to send characters
to the IBASIC Command Line field, it is permissible to set Rcv Pace and
Xmt Pace to None.
When sending data through the Test Set’s Serial Port to external
devices like printers which may have small input buffers, it is
important to set Rcv Pace and Xmt Pace to Xon/Xoff. This allows the
printer to stop data transmission from the Test Set when the printers
buffer is full and then start it again when the printer is ready.
The Test Set has a Serial Port input buffer length of 2000 characters.
Buffer size becomes important when IBASIC programs expect to
receive large amounts of data through the Serial Port with a single
ENTER statement.
Table 8-2 Test Set Serial Port 9 Configuration
Field Available Settings Recommended Setting
Serial In Inst/IBASIC Inst
IBASIC
Echo
On/Off On
Inst Echo On/Off On
Serial
Baud Rate
150, 300, 600, 1200, 2400,
4800, 9600, 19200
9,600
Parity None, Odd, Even, Always 1,
Always 0
None
Data
Length
7 bits, 8 bits 8 bits
Stop
Length
1 bit, 2 bits 1 bit
Rcv Pace
(receive
pacing)
None, Xon/Xoff Xon/Xoff
Xmt Pace
(transmit
pacing)
None, Xon/Xoff Xon/Xoff