
Chapter 1 23
Using HP-IB
Overview of the Test Set
Manual Control Mode
The Test Set’s primary instruments are shown on the left side of Figure
1-1 on page 28. There are two classes of instruments in the Test Set:
signal analyzers (RF Analyzer, AF Analyzer, Oscilloscope, Spectrum
Analyzer) and signal sources (RF Generator, AF Generator #1, AF
Generator #2).
Since so many instruments are integrated into the Test Set, it is not
feasible to have an actual “front panel” for each instrument. Therefore,
each instrument’s front panel is maintained in firmware and is
displayed on the display whenever the instrument is selected. Only one
instrument front panel can be displayed at any given time (up to four
measurement results can be displayed simultaneously if desired). Just
as with stand alone instruments, instrument front panels in the Test
Set can contain instrument setting information, measurement result(s),
or data input from the DUT (Device Under Test).
Using the Test Set in Manual Control mode is analogous to using a set
of bench or rack-mounted test equipment. To obtain a measurement
result with a bench or racked system, the desired measurement must be
“active.” For example, if an RF power meter is in the bench or racked
system and the user wishes to measure the power of an RF carrier they
must turn the power meter on, and look at the front panel to see the
measurement result. Other instruments in the system may be turned
off but this would not prevent the operator from measuring the RF
power.
Conceptually, the same is true for the Test Set. In order to make a
measurement or input data from a DUT, the desired measurement field
or data field must be “active.” This is done by using the front panel
keypad/rotary knob to select the instrument whose front panel contains
the desired measurement or data field and making sure that the
desired measurement or data field is turned ON.
Figure 1-1 on page 28 shows that instrument selection is handled by
the screen control hardware which routes the selected instrument’s
front panel to the display. Once an instrument’s front panel is displayed
on the display, the user can manipulate the instrument settings, such
as turning a specific measurement or data field on or off, using the
keypad/rotary knob. Figure 1-1 also shows that instrument setup is
handled by the Instrument Control hardware which routes setup
information from the front panel to the individual instruments.
An HP-IB/RS-232/Parallel Printer interface is located in the Test Set.
In Manual Control mode this provides the capability of connecting an
external HP-IB, serial, or parallel printer to the Test Set so that display
screens can be printed.
Kommentare zu diesen Handbüchern