16 SERVice Subsystem
730 N1911A/1912A P-Series Power Meters Programming Guide
SERVice:SENSor[1]|2:RADC?
This query returns a new raw uncorrected measurement in volts, as a 32
bit signed integer.
Syntax
Example
Error Messages
•If INIT:CONT is set to ON, error –221 “Settings Conflict” occurs.
If the E9320 Series or P- Series sensor is connected and
SERVice:SENSor[1]|2:CORRections:STATe is set to ON, error –221
“Settings Conflict” occurs.
NOTE
For E9320 Series and P-Series sensors:
Before running this query, the voltage to corrected power conversion must be disabled
using the SERVice:SENSor[1]|2:CORRections:STATe command.
SERV:SENS2:RADC? This query returns a new raw
uncorrected measurement for the sensor
connected to Channel B.
:RADC
?
SERV
1
2
:SENS