SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 395
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe
<boolean>
This command is to enable and disable the use of the currently active
sensor calibration table (CSET1) or frequency dependent offset table
(CSET2). When a table has been selected and enabled, the calibration
factors/offsets stored in it can be used by specifying the required
frequency using the [SENSe[1]]|SENSe2:FREQuency command.
When the CSET1 command is set to ON, the reference calibration factor is
taken from the sensor calibration table and is used during calibration.
Syntax
Example
Reset Condition
On reset, the sensor calibration table and frequency dependent offset table
are not affected.
CORR:CSET1:STAT 1 This command enables the use of the
currently active sensor calibration table
for Channel A.
SENS
1
:CSET
CORR
SENS2
:
:STAT
0|OFF
1|ON
Space
?
1
:CSET2