SENSe Subsystem 10
N1911A/1912A P-Series Power Meters Programming Guide 425
[SENSe[1]]|SENSe2:MRATe <character_data>
This command sets the measurement speed on the selected channel.
When a channel is set to FAST, the following couplings occur:
1
This change only occurs on the channel specified in the SENSe:MRATe command. When the
specified channel is changed from FAST to NORMal or DOUBle, the settings that were in
place when FAST was entered are restored.
2
This change occurs when either channel is set to FAST. When both channels are changed from
FAST to NORMal or DOUBle, the settings that were in place when FAST was entered are
restored.
3
Applicable to the N1912A dual channel power meter only.
Command Status
[SENSe[1]]|SENSe2:AVERage:STATe
OFF
1
[SENSe[1]]|SENSe2:CORRection:DCYCle:STATe
OFF
1
[SENSe[1]]|SENSe2:CORRection:GAIN2:STATe
OFF
1
CALCulate[1|2|3|4]:GAIN:STATe
OFF
2
CALCulate[1|2|3|4]:RELative:STATe
OFF
2
CALCulate1|3:MATH:EXPRession “(SENSe1)”
CALCulate2|4:MATH:EXPRession
“(SENSe2)”
3