Agilent Technologies 1912A Wartungshandbuch Seite 492

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 803
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 491
10 SENSe Subsystem
464 N1911A/1912A P-Series Power Meters Programming Guide
Example
Reset Condition
On reset, Y Max = 20 dBm; Y Scale = 7 dB.
Query
SENSe[1]|2:TRACe:Y:SCALe:PDIV?
The query returns the current scale setting of Y- axis in numerical value.
Query Example
Error Messages
If no sensor or wrong sensor is connected to the channel, error –241
"Hardware missing" occurs.
If channel is not in trigger acquisition mode, error –221 "Settings
conflict" occurs.
If limits of the values keyed in are exceeded, error –222 "Data out of
range; upper (or lower) limit exceeded; no change" occurs.
SENS:TRAC:Y:SCAL:PDIV
0.002
This command sets the Channel A Y Scale
value to 0.002 step.
SENS2:TRAC:Y:SCAL:PDIV
0.05
This command sets the Channel B Y Scale
value to 0.05 step.
SENS:TRAC:Y:SCAL:PDIV? This command queries the Channel A
current Y- axis scale setting in numerical
value.
Seitenansicht 491
1 2 ... 487 488 489 490 491 492 493 494 495 496 497 ... 802 803

Kommentare zu diesen Handbüchern

Keine Kommentare