Command Reference 5
Programmer’s Guide 79
Correction Apply State
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Turns On or Off the amplitude corrections. When turned On, only the
correction sets that were turned on are enabled. When turned Off, all of
the correction sets are disabled. If there is no correction enabled, state
can not be set to on
Example: :CORRection:CSET:ALL ON
Key access: Amptd > Corrections > Apply Corrections > On
Set Correction X state Off
[:SENSe]:CORRection:CSET[1]|2|3|4:OFF
[:SENSe]:CORRection:CSET[1]|2|3|4[:STATe]?
Turns the amplitude correction function on/off.
Example: :CORRection:CSET2:OFF
Key access: Amptd > Corrections > Correction1 > On
Correction Antenna Unit State
[:SENSe]:CORRection:CSET:ANTUnit OFF|ON|0|1
[:SENSe]:CORRection:CSET:ANTUnit?
Turns the antenna unit on/off.
Example: :CORRection:CSET:ANTUnit ON
Key access: Amptd > Corrections > Antenna Unit > On
Read Correction X Data
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Query amplitude correction data.
Example: :CORRection:CSET2:DATA?