Seitenansicht 124
Command Reference 5
Programmer’s Guide 117
Query OBW Result
:MEASure:OBW:OBW?
:FETCh:OBW:OBW?
Use this command to query the occupied bandwidth according to the
method you set.
Query Centroid Result
:MEASure:OBW:CENTroid?
:FETCh:OBW:CENTroid?
Use this command to query the occupied bandwidth according to the
method you set.
SEM Subsection
SEM Average Number
[:SENSe]:SEMask:AVERage:COUNt <integer>
[:SENSe]:SEMask:AVERage:COUNt?
Specifies average number in SEM measurement.
Example: :SEMask:AVERage:COUNt 20
Key access: Meas > SEM > Average Number
SEM Average State
[:SENSe]:SEMask:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:AVERage[:STATe]?
Use this command to turn on/off the average in SEM measurement.
Example: :SEMask:AVERage 1
Key access: Meas > SEM > Average Number