Agilent Technologies N9344C Wartungshandbuch Seite 137

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 224
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 136
Command Reference 5
Programmer’s Guide 129
Spectrogram Marker Time
[:SENSe]:SPECtrogram:MARKer[1]|2:TIME <value>
[:SENSe]:SPECtrogram:MARKer[1]|2:TIME?
Use this command to locate the marker to specific time point in
spectrogram measurement.
Example: :SPECtrogram:MARKer1:TIME 1 ks
Key access: Marker > Time
Spectrogram Marker Peak Search
[:SENSe]:SPECtrogram:MARKer[1]|2:PEAK
Use this command to perform Marker peak search once.
Example: :SPECtrogram:MARKer1:PEAK
Key access: Marker > Peak Search
Spectrogram Marker Previous Failure Search
[:SENSe]:SPECtrogram:MARKer[1]|2:FAIL:PRE
Use this command to locate Marker to the previous fail trace (frame) in
spectrogram measurement result.
Example: :SPECtrogram:MARKer1:FAIL:PRE
Key access: Marker > Failed Frame Search
Spectrogram Marker Next Failure Search
[:SENSe]:SPECtrogram:MARKer[1]|2:PEAK
Use this command to locate Marker to the next fail trace (frame) in
spectrogram measurement result.
Example: :SPECtrogram:MARKer1:FAIL:NEXT
Key access: Marker > Failed Frame Search