Agilent Technologies N9344C Wartungshandbuch Seite 131

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 224
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 130
Command Reference 5
Programmer’s Guide 123
SEM Offset Rel StartAmpt
[:SENSe]:SEMask:OFFSet:LIST:STARt:RCARrier
<ampl>,<ampl>,<ampl>,<ampl>,<ampl>
[:SENSe]:SEMask:OFFSet:LIST:STARt:RCARrier?
Use this command to specify the start frequency amplitude of offset
channel relative limit mask in SEM measurement. Before use this
command, the fail mask type should be set to “Relative” or “Abs AND
Rel” or “Abs OR Rel”.
Example: SENSe:SEMask:OFFSet:LIST:STARt:RCARrier -14,
-14, -15,-20, -20
Key access: Meas > SEM > Offset/Limit > More 1 of 3 > Rel StratAmpt
SEM Offset Abs Stop Ampt
[:SENSe]:SEMask:OFFSet:LIST:STOP:ABSolute
<ampl>,<ampl>,<ampl>,<ampl>,<ampl>
[:SENSe]:SEMask:OFFSet:LIST:STOP:ABSolute?
Use this command to specify the stop frequency amplitude of offset
channel absolute limit mask in SEM measurement. Before use this
command, the fail mask type should be set to “Absolute” or “Abs AND
Rel” or “Abs OR Rel”.
Example: :SENSe:SEMask:OFFSet:LIST:STOP:ABSolute -10,
-20, -20, -30, -30
Key access: Meas > SEM > Offset/Limit > More 1 of 3 > Abs Stop Ampt
SEM Offset Rel StopAmpt
[:SENSe]:SEMask:OFFSet:LIST:STOP:RCARrier
<ampl>,<ampl>,<ampl>,<ampl>,<ampl>
[:SENSe]:SEMask:OFFSet:LIST:STOP:RCARrier?
This command defines the stop frequency amplitude of offset channel
relative limit mask in SEM measurement. The fail mask type should be
set to “Relative” or “Abs AND Rel” or “Abs OR Rel”.
Example: SENSe:SEMask:OFFSet:LIST:STOP:RCARrier -10, -20,
-20, -30, -30
Key access: Meas > SEM > Offset/Limit > More 1 of 3 > Rel Stop Ampt
Seitenansicht 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 223 224

Kommentare zu diesen Handbüchern

Keine Kommentare