Mask Test 11
InfiniiVision Oscilloscopes User’s Guide 339
Mask Statistics
From the Mask Test Menu, press the Statistics softkey to enter the Mask Statistics
Menu.
When you enable Show Statistics the following information is displayed:
Current mask, name of mask, Channel number, date and time.
# of Tests (total number of mask tests executed).
Status (Passing, Failing, or Untested).
Accumulated test time (in hours, minutes, seconds, and tenths of seconds).
And for each analog channel:
Number of failures (acquisitions in which the signal excursion went beyond the mask).
Failure rate (percentage of failures).
Sigma (the ratio of process sigma to maximum achievable sigma, based on number of
waveforms tested).