Agilent Technologies E8364C Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messen, Testen & Steuern Agilent Technologies E8364C herunter. Agilent E8364C Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Technical Specifications

Technical SpecificationsAgilent TechnologiesPNA Series Network AnalyzersE8362B/C, E8363B/C, and E8364B/CManufacturing Part Number: E8364-90031Printed

Seite 2 - Documentation Warranty

6 the analyzer can have predefined frequency segments which will output a higher power level when the extended dynamic range is required (i.e. device

Seite 3 - Contacting Agilent

96 Table 32. X11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 4

97NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 5 - (Rev. 2008-10-03)

98 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of

Seite 6

99Table 33. Uncorrected System Performancea Specifications apply over environmental temperature of 23° ±3 °C, with < 1 °C deviation from the calib

Seite 7

100 Table 33 (Continued). Uncorrected System Performancea Reflection Tracking Typical: 10 MHz to 45 MHzb -- ±1.5 dB 45 MHz to 20 GHz -- ±

Seite 8 - Corrected System Performance

101Table 33 (Continued). Uncorrected System Performancea Crosstalk - Option 080 enabledb,e Typical: 10 MHz to 45 MHz -- -65 45 MHz to 1 GHz

Seite 9 - Table 1. System Dynamic Range

102 Table 34. Test Port Output Description Specification Supplemental Frequency Range Standard Opt 014 Opt UNL Opt 014/UNL E8362B/C 10 MHz

Seite 10

103Table 34 (Continued). Test Port Output Power Rangea, e, f 10 MHz to 45 MHzb -25 to +2 dBm (typical) -25 to +2 dBm (typical)-87 to +2 dBm (typi

Seite 11

104 Table 34 (Continued). Test Port Output 10 kHz offset from center frequency, nominal power at test port Typical: 10 MHz to 45 MHz -

Seite 12

105Table 34 (Continued). Test Port Output 1 MHz offset from center frequency, nominal power at test port - Option 080 enabled Typical:

Seite 13 - (E8363B/C AND E8364B/C)

7Table 2. Receiver Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Dynamic Range (in a 1

Seite 14

106 Table 35: Test Port Input Description Specification Supplemental Standard Opt 014 Opt UNL Opt 014/UNL Test Port Noise Floora 10 Hz IF Ba

Seite 15

107Table 35 (Continued). Test Port Input Test Port Noise Floora,b Option 080 enabledd 10 Hz IF Bandwidth 10 MHz to 45 MHzb <-77 dBm (typical) <

Seite 16

108 Table 35 (Continued). Test Port Input Direct Receiver Access Input Noise Floora,b 10 Hz IF Bandwidth 10 MHz to 45 MHz -- <-127 dBm -- &l

Seite 17

109Table 35 (Continued). Test Port Input Direct Receiver Access Input Noise Floora,b - Option 080 enabledd 10 Hz IF Bandwidth 10 MHz to 45 MHz -- &

Seite 18

110 Table 35 (Continued). Test Port Input Third Order Intercept - Tone spacing from 100 kHz - 5 MHz Typical 10 MHz to 150 MHz -- +33 dBm 150

Seite 19

111Table 35 (Continued). Test Port Input Trace Noise Magnitude - Option 080 enabledb,d 1 kHz IF bandwidth. Ratio measurement, nominal power at test p

Seite 20

112 Table 35 (Continued). Test Port Input Stability Magnituded Typical ratio measurement, made at the test port. 10 MHz to 45 MHz -- -- -- --

Seite 21

113Table 36. Dynamic Accuracy (Specificationa) Accuracy of the test port input power reading relative to the reference input power level. NOTE: The

Seite 24

8 Table 2 (Continued). Receiver Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Dynamic

Seite 25

116 a Dynamic accuracy is verified with the following measurements:  Compression over frequency.  IF linearity at a single frequency of 1.

Seite 26

117Table 37. Test Port Input (Group Delay)a Description Specification Supplemental Information (typ.) Aperture (selectable) (frequency span)/(num

Seite 27

118 General Information Table 38. Miscellaneous Information Description Specification Supplemental Information System IF Bandwidth Range -- 1 Hz

Seite 28

119Table 40. Rear Panel Information Description Supplemental Information 10 MHz Reference In Connector BNC, female Input Frequency 10 MHz ± 10 ppm

Seite 29

120 Table 40 (Continued). Rear Panel Information Test Set Drivers Function Used for driving remote mixers Connectors SMA (f) for RF and LO outputs

Seite 30

121Table 40 (Continued). Rear Panel Information USB Port One port on front panel and five ports on rear panel. Universal Serial Bus jack, Type A c

Seite 31 - (E836xB/C)

122 Table 41. Analyzer Dimensions and Weight Description Supplemental Information Cabinet Dimensions Height Width Depth Excluding front and rear

Seite 32

123Measurement Throughput Summary Table 42 Typical Cycle Timea,b (ms) for Measurement Completion Number of Points 201 401 1601 16,001 Start 28 G

Seite 33

124 Table 43. Cycle Time vs IF Bandwidtha Applies to the Preset condition (201 points, correction off) except for the following changes:  CF = 28 G

Seite 34

125Table 45. Frequency Converter Application (option 083) Cycle Time for Fixed-IF Measurements (ms)1 Number of Points 101 201 401 Stimulus star

Seite 35

9E8363B/C AND E8364B/C Corrected System Performance with 2.4mm Connectors ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Table 3. 85056A Calibration Ki

Seite 36

126 Specifications: Front-Panel Jumpers ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Models E8362B/C, E8363B/C, and E8364B/C Option 014 ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯

Seite 37

127 E8363B/C: 10 Hz IF Bandwidth 45 MHz to 500 MHz < -127 dBm -- 500 MHz to 2 GHz < -133 dBm -- 2 GHz to 10 GHz <

Seite 38

128 Table 48: Reference Receiver Inputs (Rcvr R1, Rcvr R2) Description Specification Supplemental Information Minimum Input Level E8362B/C, E8363

Seite 39

129 Table 49: Reference Outputs (Reference 1 Source Out, Reference 2 Source Out) Description Specification Supplemental Information Maximum Output

Seite 40

130 Table 50: Source Outputs (Port 1 Source Out, Port 2 Source Out) Description Specification Supplemental Information Maximum Output Level E8362

Seite 41

131 E8364B/C, Option 014 and UNL: 45 MHz to 500 MHz -- 3.5 dBm, typical 500 MHz to 2 GHz -- 5 dBm, typical 2 GHz to 10 GHz --

Seite 42

132 Table 51: Coupler Inputs (Port 1 Cplr Thru, Port 2 Cplr Thru) Description Specification Supplemental Information Insertion Loss to Test Port

Seite 43

133 E8364B/C, Option 014 and UNL: 45 MHz to 500 MHz -- 0.5 dB, typical 500 MHz to 2 GHz -- 1 dB, typical 2 GHz to 10 GHz -- 2

Seite 44

134 Test Set Block Diagrams E836xB/C - Standard Configuration and Standard Power Range E836xB/C - Option UNL Standard Configuration with Extende

Seite 45

135E836xB/C - Option UNL Standard Configuration with Extended Power Range and Bias - Tees, and Option 016, Receiver Attenuators

Seite 46

10 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 47

136 Test Set with Option 014 Block Diagrams E836xB/C - Option 014 Configurable Test Set and Standard Power Range Item Description Item Descript

Seite 48

137E836xB/C - Option 014 Configurable Test Set and Standard Power Range, and Option 081 Reference Channel Transfer Switch Item Description Item

Seite 49

138 E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees Item Description Item Description a SOURCE

Seite 50

139E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees, and Option 081 Reference Channel Transfer Switch

Seite 51

140 E836xB/C - Option 014 Configurable Test Set and Option UNL, Extended Power Range and Bias - Tees and Option 016 Receiver Attenuators Item Desc

Seite 52

141E836xB/C - Option 014 Configurable Test Set, and Option UNL Extended Power Range and Bias - Tees, and Option 016 Receiver Attenuators, and Option

Seite 53

11 Reflection Uncertainty (Specifications)

Seite 54

12 Table 4. 85056A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 55

13NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 56

14 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of

Seite 57

15Table 5. 85056D Calibration Kit Standard Configuration and Standard Power Range (E8363B/C AND E8364B/C) Applies to the, E8363B/C AND E8364B/C an

Seite 58

Documentation WarrantyTHE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTU

Seite 59

16 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 60

17 Reflection Uncertainty (Specifications)

Seite 61

18 Table 6. 85056D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 62

19NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E

Seite 63

20 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offs

Seite 64

21E8363B/C AND E8364B/C Corrected System Performance with 2.92mm Connectors Table 7. 85056K Calibration Kit Standard Configuration and Standard Powe

Seite 65

22 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 66

23 Reflection Uncertainty (Specifications)

Seite 67

24 Table 8. 85056K Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 68

25 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications * Configurable Test Set,

Seite 69

Contacting AgilentAssistance with test and measurement needs and information on finding a local Agilent office are available on the Web at:www.agilent

Seite 70

26 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offse

Seite 71

27E836xB/C Corrected System Performance with 3.5mm Connectors ⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯⎯ Table 9. 85052B Calibration Kit Standard

Seite 72

28 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 73

29 Reflection Uncertainty (Specifications)

Seite 74

30 Table 10. 85052B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 75

31NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 76

32 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of

Seite 77

33Table 11. 85052C Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85052C (3.5mm) c

Seite 78

34 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 79

35 Reflection Uncertainty (Specifications)

Seite 80

This page intentionally left blank.

Seite 81

36 Table 12. 85052C Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 82

37NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 83

38 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency O

Seite 84

39Table 13. 85052D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85052D (3.5mm) c

Seite 85

40 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 86

41 Reflection Uncertainty (Specifications)

Seite 87

42 Table 14. 85052D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 88

43NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 89

44 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency O

Seite 90

45E836xB/C Corrected System Performance with 7mm Connectors Table 15. 85050B Calibration Kit Standard Configuration and Standard Power Range (E836

Seite 91

1Technical Specifications for the E8362B/C, E8363B/C, E8364B/C (Rev. 2008-10-03) Definitions...

Seite 92

46 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 93

47 Reflection Uncertainty (Specifications)

Seite 94

48 Table 16. 85050B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 95

49NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set, E

Seite 96

50 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency

Seite 97

51Table 17. 85050C Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85050C (7mm) cal

Seite 98

52 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 99

53 Reflection Uncertainty (Specifications)

Seite 100 - Transfer Switch

54 Table 18. 85050C Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 101

55NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set,

Seite 102

2 Standard Power Range (E836xB/C)... 63 Table 22. 8505

Seite 103

56 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency

Seite 104

57Table 19. 85050D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85050D (7mm) cal

Seite 105

58 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 106 - Table 34. Test Port Output

59 Reflection Uncertainty (Specifications)

Seite 107

60 Table 20. 85050D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 108

61NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) *Configurable Test Set

Seite 109

62 Reflection Uncertainty (Specifications) *Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequen

Seite 110 - Table 35: Test Port Input

63E836xB/C Corrected System Performance with Type-N Connectors Table 21. 85054B Calibration Kit Standard Configuration and Standard Power Range (E

Seite 111

64 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 112

65 Reflection Uncertainty (Specifications)

Seite 113

3Table 48: Reference Receiver Inputs (Rcvr R1, Rcvr R2) ... 128 Table 49: Reference Outputs

Seite 114

66 Table 22. 85054B Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 115

67NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 116

68 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency

Seite 117

69Table 23. 85054D Calibration Kit Standard Configuration and Standard Power Range (E836xB/C) Applies to the, E836xB/C analyzers, 85054D (Type-N)

Seite 118

70 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 119

71 Reflection Uncertainty (Specifications)

Seite 120 - Compression over frequency

72 Table 24. 85054D Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 121

73NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E

Seite 122 - General Information

74 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of

Seite 123

75E8363B/C AND E8364B/C Corrected System Performance with WR-28 Connectors Table 25. R11644A Calibration Kit Standard Configuration and Standard Pow

Seite 124

4 Definitions All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument ha

Seite 125

76 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 126

77 Reflection Uncertainty (Specifications)

Seite 127 - Table 42 Typical Cycle Time

78 Table 26. R11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 128

79NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set, E

Seite 129 - Typical performance

80 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Offs

Seite 130

81E8363B/C AND E8364B/C Corrected System Performance with WR-42 Connectors Table 27. K11644A Calibration Kit Standard Configuration and Standard Pow

Seite 131

82 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 132

83 Reflection Uncertainty (Specifications)

Seite 133

84 Table 28. K11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 134

85NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 135

5Table 1. System Dynamic Rangea Description Specification (dB) at Test Portb Typical (dB) at Direct Receiver Access Inputc Supplemental InformationD

Seite 136

86 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequency Of

Seite 137

87E836xB/C Corrected System Performance with WR-62 Connectors Table 29. P11644A Calibration Kit Standard Configuration and Standard Power Range (E

Seite 138

88 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 139 - Receiver Attenuators

89 Reflection Uncertainty (Specifications)

Seite 140

90 Table 30. P11644A Calibration Kit Fully Optioned (E836xB/C - Option 014, UNL, 016, 080, and 081) Configurable Test Set, Extended Power Range &

Seite 141

91NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications) * Configurable Test Set,

Seite 142

92 Reflection Uncertainty (Specifications) * Configurable Test Set, Extended Power Range & Bias-Tees, Receiver Attenuators, Frequenc

Seite 143

93E836xB/C Corrected System Performance with WR-90 Connectors Table 31. X11644A Calibration Kit Standard Configuration and Standard Power Range (E

Seite 144

94 NOTE: The following graphs also apply to the “C” model of the analyzers. Transmission Uncertainty (Specifications)

Seite 145

95 Reflection Uncertainty (Specifications)

Kommentare zu diesen Handbüchern

Keine Kommentare