Table 11. 85052C Calibration Kit
Standard Configuration and Standard Power Range
(E836xB/C)
Applies to the, E836xB/C analyzers, 85052C (3.5mm) calibration kit, 85131F flexible test port cable set, and a full 2-port calibration. Data
and traces above 20 GHz are not applicable to the E8362C. Also applies to the following condition:
Environmental temperature 23° ±3 °C, with < 1 °C deviation from calibration temperature
Specification (dB) Description
45 MHz to
2 GHz
2 to
20 GHz
20 to
26.5 GHz
Directivity 48 50 50
Source Match 40 50 50
Load Match 48 50 50
Reflection Tracking
±0.003
+0.02/°C
±0.000
+0.02/°C
±0.000
+0.03/°C
Transmission Tracking
±0.009
+0.02/°C
±0.014
+0.02/°C
±0.018
+0.03/°C