7
Measurement Specifications
Range
Measurement Resolution
Accuracy
(% reading + offset)
B2901A/B2902A B2911A/B2912A
±200 mV 100 nV 100 nV ±(0.015 % + 225 μV)
±2 V 1 μV 1 μV ±(0.02 % + 350 μV)
±20 V 10 μV 10 μV ±(0.015 % + 5 mV)
±200 V 100 μV 100 μV ±(0.015 % + 50 mV)
1. Supplemental characteristics
Range
Measurement Resolution
Accuracy
(% reading + offset)
B2901A/B2902A B2911A/B2912A
±10 nA
1
--- 10 fA ±(0.10 % + 50 pA)
±100 nA 100 fA 100 fA ±(0.06 % + 100 pA)
±1 μA 1 pA 1 pA ±(0.025 % + 500 pA)
±10 μA 10 pA 10 pA ±(0.025 % + 1.5 nA)
±100 μA 100 pA 100 pA ±(0.02 % + 25 nA)
±1 mA 1 nA 1 nA ±(0.02 % + 200 nA)
±10 mA 10 nA 10 nA ±(0.02 % + 2.5 μA)
±100 mA 100 nA 100 nA ±(0.02 % + 20 μA)
±1 A 1 μA 1 μA ±(0.03 % + 1.5 mA)
±1.5 A 1 μA 1 μA ±(0.05 % + 3.5 mA)
±3 A 10 μA 10 μA ±(0.4 % + 7 mA)
±10 A
2
10 μA 10 μA ±(0.4 % + 25 mA)
3
Current Measurement Specifications
1.10 nA range is not available for B2901A/B2902A.
2.10 A range is available only for pulse mode, not available for DC mode.
3.Measurement speed: 0.01 PLC
Voltage Measurement Specifications