7
Sweep Capabilities
Sweep mode: linear, logarithmic (log) or list
Sweep direction: single or double
Type: DC, or pulse
Number of steps: 1 to 100,000
Min programmable value to create list sweep waveform:
Minimum 10 μs with 1 μs resolution
Arbitrary Waveform Generation Capabilities
Pre-defined waveforms and user-defined waveforms are supported.
Pre-defined waveforms:
Supported waveforms: Sine, Square, Ramp, Triangle, Trapezoid and Exponential
Frequency: 1 mHz to 10 kHz
Programmable frequency/timing resolution: 1 μHz nominal (sine), 250 ns
(waveforms other than sine)
Frequency accuracy of time base: ±50 ppm
User-defined waveforms:
Waveform length: 1 to 100,000 points
Sample rate: 0.001 to 100,000 Sa/s, 250 ns resolution
Storage: Non-volatile memory and USB memory are both available.
Non-volatile memory can store one waveform with a length of up to 2500
points. USB memory can store waveforms of up to 100k points.