3
Specification conditions
The source and measurement accuracy are specified under the conditions
listed below.
1. Temperature: 23 °C ± 5 °C
2. Humidity: 30 % to 80 % RH
3. After 60 minutes warm-up
4. Ambient temperature change less than ±3 °C after self-calibration execution
5. Calibration period: 1 year
6. Measurement speed: 1 PLC (Power Line Cycle)
Specifications
Maximum voltage and current
Max Voltage Max Current
DC or Pulsed
1
210 V 0.105 A
21 V 1.515 A
2
6 V 3.03 A
2
Pulsed only
1
200 V 1.515 A
6 V 10.5 A
1. See “Maximum pulse width and duty cycle” in Pulse Source Supplemental
Characteristics for applicable maximum voltage and current.
2. Max current limitation: For 21 V/1.515 A and 6 V / 3.03 A ranges, total max current is
limited by the table below for using 2 channels. Max current is not limited for using
1 channel only.
Maximum current limitation
-200
Current [A]
Voltage [V]
Pulsed only
1
-10.5
-6 6
200
1.515
10.5
-1.515
3.03
Voltage [V]
21
-21
-6
-0.1
6
-1.515
0.1
1.515
-3.03
-210
210
DC or Pulsed
1
Current [A]
Ch1 voltage Ch2 voltage Max total current limitation of Ch1 and Ch2
±(0 V< V ≤6 V) ±(0 V< V ≤6 V) Ch1 current + Ch2 current ≤ 4 A
±(0 V< V ≤6 V) ±(6 V < V ≤21 V) Ch1 current + Ch2 current x 1.6 ≤ 4 A
±(6 V < V ≤21 V) ±(0 V< V ≤6 V) Ch1 current + Ch2 current x 0.625 ≤ 2.5 A
±(6 V < V ≤21 V) ±(6 V < V ≤21 V) Ch1 current + Ch2 current ≤ 2.5 A