Agilent B1500A/B1505A Programming Guide, Edition 11 1-65
Programming Basics
Programming Tips
To Use Programs for Agilent 4155/4156
Agilent B1500 supports commands similar to the FLEX command of the Agilent
4155B/4156B/4155C/4156C Parameter Analyzer. However, not all command sets
are fully compatible. To reuse the programs created for the Agilent 4155/4156, the
following modifications are required.
• To remove all unsupported commands
Table 1-18 shows the commands not supported by the B1500. You cannot use
these commands. The SCPI commands and 4145 syntax commands are not
supported neither.
The B1500 does not need the US and :PAGE commands that are necessary to
change the control mode of the Agilent 4155/4156.
• To check and correct the command syntax
Even if the command name is the same, the available parameters and values may
be different. Check and correct the command parameters.
• To change the FMT command parameter
Use the FMT 21, FMT 22, or FMT 25 command that sets the data output format
compatible with the 4155/4156 ASCII format.
• To delete RMD?
The B1500 does not need the RMD? command that is necessary to put the
measurement data into the output data buffer of the Agilent 4155/4156.
• FL command
The initial setting of the FL command is different. It is ON for the Agilent
4155/4156, and OFF for the B1500.
Add the FL1 command to use the filter.
• AV command
This command is used to set the A/D converter of the B1500.
To set the high resolution ADC installed in the B1500, use the AAD and AIT
commands.
• To replace TI?/TV?/TTI?/TTV? with TI/TV/TTI/TTV respectively
• To replace WM with LSM for the linear search measurement
• To replace TSQ? with TSQ