4-66 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
CORRL?
Example
Statements
OUTPUT @B1500;"CORRL 9,3000000"
CORRL?
This command returns the frequency stored in the frequency list for the MFCMU
correction data measurement.
Syntax CORRL? chnum[,index]
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
index : Index number of the list. Integer expression.
Query Response CORRL? chnum returns:
number_of_frequencies<CR/LF^EOI>
This value is the number of frequencies stored in the list.
CORRL? chnum,index returns:
frequency<CR/LF^EOI>
This value is the frequency corresponding to the specified index.
Example
Statements
OUTPUT @B1500;"CLCORR 9,2"
OUTPUT @B1500;"CORRL? 9"
ENTER @B1500;Number
OUTPUT @B1500;"CORRL? 9,4"
ENTER @B1500;Freq
This example returns Number=12 and Freq=10000.
CORRSER?
This query command measures the SPGU terminal voltage and returns the
measurement data and the calculated impedance data. Setting mode=1 automatically
executes the SER command using this impedance data as the input parameter. Only
one channel can be specified in this command. So the voltage monitor cannot be
performed by using multiple channels simultaneously.
Execution
Conditions
The SPGU output channel must be set up by using the following commands before
executing this command.
SPPER, SPM, SPT, and SPV commands, in PG mode
ALW and ALS commands, in ALWG mode