Seitenansicht 267
3-90 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
CV (AC Level) Sweep Measurements
CV (AC Level) Sweep Measurements
To perform capacitance-voltage (AC level) sweep measurements, use the following
commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets SMU filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets SMU series resistor ON/OFF [SSR] chnum,mode
Disables SCUU status indicator [SSL] chnum,mode
Controls SCUU input-output path [SSP] chnum,path
Sets MFCMU A/D converter [ACT] mode[,N]
Sets MFCMU measurement mode [IMP] mode
Sets AC/DC voltage monitor ON/OFF [LMN] mode
Sets MFCMU output frequency FC chnum,freq
Forces AC voltage by using MFCMU ACV chnum,level
Sets CV sweep timing parameter WTACV hold,delay
[,sdelay[,tdelay[,mdelay]]]
Sets auto abort function [WMACV] abort[,post]
Sets AC level sweep source WACV chnum,mode,start,stop,step
Forces DC voltage by using MFCMU DCV chnum,voltage
Forces constant voltage [DV, TDV] chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current [DI, TDI]
Sets MFCMU measurement range [RC] chnum,mode
[,range]
Sets measurement mode MM 23,chnum
Executes measurement XE