Agilent B1500A/B1505A Programming Guide, Edition 11 2-23
Remote Mode Functions
Measurement Modes
NOTE The LSM command sets the automatic abort function and the output after search.
The LSTM command sets the hold time and delay time. The LSV/LSI command
sets the search output, and the LGI/LGV command sets the measurement channel.
To Use
Synchronous
Output Channel
You can use the synchronous output channel that provides output synchronized with
the search source. Refer to “Synchronous Output” on page 2-44.
After measurement, the synchronous channel forces the start+offset, stop+offset, or
the last output value, as specified by the LSM command, and keeps it.
Figure 2-15 Synchronous Output
NOTE The LSSV/LSSI command sets the synchronous output. You can use the same
output mode (voltage or current) as the search source. All output values must be
covered by the output range of the search source.
Voltage or current
Start value
Time
Stop value
Primary search source
Synchronous search source
Previous value
Trigger (e.g. XE)
LSM, LSTM, and LSV/LSSV or LSI/LSSI
Offset
Offset
Offset