Frequency Sweep Commands 8
U2761A Programmer’s Reference 91
[SOURce:]FREQuency:STOP
Syntax
[SOURce:]FREQuency:STOP <frequency>
This command sets the stop frequency.
[SOURce:]FREQuency:STOP?
This query returns the stop frequency in Hz.
Parameter
Examples
The following command sets the stop frequency to 10 kHz.
FREQ:STOP 10000
The following query returns the stop frequency.
FREQ:STOP?
Typical response:
+1.0000000000000E+04
Item Type Range of values Default value
frequency Numeric 1 μHz to 20 MHz (for sine and square)
1 μHz to 200 kHz (for ramp and arbitrary)
1 kHz
SOUR : FREQ space
?
frequency:STOP