CALibration Commands 14
U2761A Programmer’s Reference 135
CALibration:STRing
Syntax
CALibration:STRing <quoted string>
This command stores a message in nonvolatile calibration memory. Storing
a message will overwrite any message previously stored in memory.
CALibration:STRing?
This query reads the calibration message and returns a quoted string.
Parameter
Examples
The following command stores the calibration message.
CAL:STR '01 JAN 2008:CALIBRATED'
The following query returns the quoted string.
CAL:STR?
Typical response:
''01 JAN 2008:CALIBRATED''
Item Type Range of values Default value
string String Up to 40 alphanumeric characters N/A
CAL space
?
quoted string:STR