Seitenansicht 66
Pulse Configuration Commands 5
U2761A Programmer’s Reference 59
[SOURce:]FUNCtion:PULSe:HOLD
Syntax
[SOURce:]FUNCtion:PULSe:HOLD {WIDTh|DCYCle}
This command sets the U2761A to hold either the pulse width or pulse
duty cycle setting constant as the period is varied.
[SOURce:]FUNCtion:PULSe:HOLD?
The query returns the pulse hold setting as WIDT or DCYC.
Parameter
Examples
The following command sets the U2761A to hold the pulse width.
FUNC:PULS:HOLD WIDT
The following query returns the pulse hold setting.
FUNC:PULS:HOLD?
Typical response:
WIDT
Item Type Range of values Default value
hold Discrete WIDTh|DCYCle WIDTh
SOUR : FUNC space
?
:PULS :HOLD WIDT
DCYC