
Chapter 2 Service
The Agilent 53181A Calibration Procedures
2-32 Assembly-Level Service Guide
2
c. Press key once, and then press key three times.
The Counter should display CODE: 53000.
d. Press key once, and then press key once.
The Counter should display CODE: 53100.
e. Press key once, and then press key eight times.
The Counter should display CODE: 53180.
f. Press key once, and then press key once.
The Counter should display CODE: 53181.
g. Finally, press Enter key.
CAL UNSECURE is displayed.
NOTE If “53181” is NOT the correct security code, the Counter will display BAD CODE.
4 Press Scale & Offset key until CAL: OFFS 1 ? is displayed.
5 Press Enter key.
A scrolling message on the display will ask you to remove any signals connected to
Channel 1. Follow the instruction.
6 Press Enter key.
The Counter momentarily displays CALIBRATING, and then it should display
OFFS 1 PASS.
If the fail message is displayed, refer to the troubleshooting section in this chapter.
This completes the Offset Calibration procedure. Proceed to the next procedure to
calibrate the gain.
Kommentare zu diesen Handbüchern