Agilent Technologies HP 3325B Bedienerhandbuch Seite 121

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 252
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 120
Chapter 2 Service
The Agilent 53131A/132A Calibration Procedures
Assembly-Level Service Guide 2-23
2
To Calibrate the Standard Timebase
1 Connect the output of a 10 MHz house standard to Channel 1 of the Counter as
shown in Figure 2-5.
Figure 2-5. Standard Timebase Calibration Setup
2Cycle the POWER key to preset the Counter.
Allow the counter to warm up for approximately 30 minutes.
3 With an insulated tuning tool, turn the OSC Adjust potentiometer (located on the
rear chassis of the Counter) until the frequency reading in the Counter display is
10.00000000 MHz ± 5 Hz.
The Counter will be calibrated to the house standard.
If you are not able to adjust the frequency correctly, refer to the troubleshooting
section in this chapter.
This completes the Standard Timebase Calibration procedure.
H
P 53131A/132A
C
ounter
Channel 1
House Standard
10 MHz Reference
Seitenansicht 120
1 2 ... 116 117 118 119 120 121 122 123 124 125 126 ... 251 252

Kommentare zu diesen Handbüchern

Keine Kommentare