Agilent Technologies 6834B Betriebsanweisung Seite 213

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 243
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 212
M3000A/M3046AM3015A/M3016A Service Guide
Repairing the Instrument 6-13
Disassembly Procedures for the Measurement Server Extension
It is recommended that you replace all the replaceable parts in the Extension (CO
2
Scrubber,
Infrared Lamp and Pump) after 15 000 hours (approximately 3 years) of continuous use. See
Chapter 3 Maintaining the Instrument for more details on caring for the Extension.
7RROV5HT XLUHG
A thin-bladed screwdriver.
A pair of large tweezers.
In addition, for removing the Pump, you will need a large-bladed screwdriver.
:$51,1*
7KHUHLVKLJKYROWDJHLQVLGHWKH,QVWUXPHQW9'RQRWFRQQHFWWKH0HDVXUHPHQW
6HUYHU([WHQVLRQWRD0RQLWRUZKLOHWKH([WHQVLRQKRXVLQJLVRSHQ
$VZHOOSDUWVLQVLGHWKH,QVWUXPHQWPD\EHFRQWDPLQDWHGZLWKEDFWHULD3URWHFW
\RXUVHOIIURPSRVVLEOHLQIHFWLRQE\ZHDULQJH[DPLQDWLRQJORYHVGXULQJWKHVH
SURFHGXUHV
5HPRYLQJWKH)URQW&RYHU
To remove the front cover, do the following:
6WHS Remove the Server and the Monitor from the Extension.
6WHS Use a thin-bladed screwdriver to prise the grey front cover (the console covering
the measurement connector hardware) gently from the bottom of the Extension.
Position the screwdriver in the small slits provided for this purpose. The front cover
then clicks away from the Extension.
6WHS Remove the front cover.
5HPRYLQJWKH([WHQVLRQ%RWWRP&RYHU
To remove the Extension bottom cover, do the following:
Seitenansicht 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 242 243

Kommentare zu diesen Handbüchern

Keine Kommentare