CAUTION
Store the electrode cartridge in the provided box. If the pins of the electrode cartridge
are bent or misaligned, poor quality results or pre-terminated chip runs will result.
Insert the pressure cartridge:
1 Slide the pressure cartridge in the lid as shown below.
2 Push the metal front of the cartridge to ensure a tight connection.
3 Push the metal locking lever in the flat (closed) position.
CAUTION
Do not force the lid closed. This may damage the cartridge. If the lid does not close
completely, check that the cartridge is inserted properly and try again to close the lid.
Metal lever
Push here to ensure
tight connection