Troubleshooting the Logic Analyzer
Inverse Assembler Problems
Inverse assembler will not load or run
You need to ensure that you have the correct system software loaded
on your analyzer.
Ensure that the inverse assembler is on the same disk as the configuration
files you are loading.
Configuration files for the state analyzer contain a pointer to the location
of the corresponding inverse assembler. If you delete the inverse
assembler or move it to another location, the configuration process will
fail.
Make sure you are using the version of the inverse assembler software that
corresponds to the operating system revision installed on your analyzer.
See your Analysis Probe User's Guide for details.