Agilent Technologies DP235 Betriebsanweisung Seite 455

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 628
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 454
455
Operator’s Service
Preparing For Use
Preparing For Use
This section gives you instructions for preparing the logic analyzer for
use.
Power requirements
The logic analyzer requires a power source of either 115 VAC or 230
VAC, -22 % to +10%, single phase, 48 to 66 Hz, 200 Watts maximum
power.
Category I: Signal level, telecommunications, electronic. (Seperated
from line voltage by transformer, etc).
Category II: Local level, appliances, portable equipment (line voltage in
appliance and to wall outlet).
Operating environment
The operating environment is listed in previous chapter. Note the
noncondensing humidity limitation. Condensation within the
instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
The logic analyzer will operate at all specifications within the
temperature and humidity range given. However, reliability is
enhanced when operating the logic analyzer within the following
ranges:
Temperature: +20°C to +35 °C (+68 °F to +95 °F)
Humidity: 20% to 80% noncondensing
Pollution degree 2: Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be
expected.
Seitenansicht 454
1 2 ... 450 451 452 453 454 455 456 457 458 459 460 ... 627 628

Kommentare zu diesen Handbüchern

Keine Kommentare