Operating the 8000 4
8000 Operator’s Manual 53
Main Menu Option 2 - Set DISSO
From the Main Menu, select option 2, Set DISSO. Enter a program
number and press ENTER. The DISSO Parameters screen displays.
Following is a description of the DISSO Parameters screen options:
***DISSO PARAMETERS*** 4 SET BATH TEMP (VESSEL TEMP)
1 SET SPINDLE RPM 5 COM PORT ID
2 FINAL SPIN LENGTH 6 DDM OR BASKET
3 INI/FINAL VESSEL TEMP 7 PRINT FREQ
DISSO applies to all dissolution apparatus.
Table 5 Disso Parameters Options (Apparatus 1 / 2)
Option Function
1 Set Spindle RPM Select option 1, Set Spindle RPM. The Parameter for Manual or Program screen
displays.Select option 1, Manual, to set an RPM to be used as the default for manual
operation (see “Main Menu Option 6 - Control DISSO” on page 57). Optionally,
select option 2, Program, to set an RPM to be used as a default for programmed
operation (see “Starting a 8000 Program” on page 66). Enter the desired RPM and
press ENTER. The DISSO Parameters screen displays.
2 Final Spin Length Select option 2, Final Spin Length, to program a short, high-speed spin cycle at
250 RPM at the end of the test. Enter a duration for the final spin and press ENTER.
The DISSO Parameters screen displays.
Note: Many dissolution protocols now call for an infinity sample to be taken after the
end of the test.
3 Ini/Final Vessel Temp Select option 3, Ini/Final Vessel Temp. The Initial Vessel Temp Measurement screen
displays. Select option 1 to enable or option 2 to disable the initial vessel temperature
reading. The Final Vessel Temp Measurement screen displays. Select option 1 to
enable or option 2 to disable the final vessel temperature reading. The DISSO
Parameters screen displays.