100 8000 Operator’s Manual
4 Operating the 8000
15 Upon completion of the priming cycle, the following screen displays:
16 Press OPEN VALVES. The first calibration tube begins filling.
17 Observe the first calibration tube and press OPEN VALVES again
when 10 mL is reached.
18 Repeat Steps 17 and 18 for the remaining calibration tubes. When the
last valve is calibrated, the pump reverses and the lines are purged.
The calibration completion alarm, if enabled, sounds when the
calibration sequence is finished.
19 Press any key to silence the alarm.
20 After the dispensing arm moves forward, remove and clean the
calibration tubes.
21 Reconnect the top and bottom portions of the autocalibration block.
22 Press ESC.
23 Press MENU.
24 Select option 7, Manual Operation.
25 Select option 4, Goto Rows.
26 Enter 4 to move the dispensing arm to row 4.
27 Replace the media rinse reservoir in front of the autocalibration
block.
28 Press H to return to the home position.
29 Press ESC three times to return to the Ready screen.
PRESS <OPEN VALVES> TO START
PRESS AGAIN WHEN 10 ML IS REACHED
MANUAL CALIBRATION OF VALVE # 1
TIME: 01/01/12 21:56:40 ELAPSE: 000:00:00
NOTE
If you press ESC before all the valves are calibrated, you cancel the
calibration and return to the previously saved values for each valve.
None of the new calibration results are saved. The calibration must be
allowed to finish before the new calibration values are saved.