Setting Up the 8000 3
8000 Operator’s Manual 37
If your system configuration includes the peristaltic pump, complete the
steps under “Installing and Connecting the Autocalibration Block
(Peristaltic Pump only)” on page 37.
Installing and Connecting the Autocalibration Block
(Peristaltic Pump only)
1 With the dispensing arm forward, insert the autocalibration block
behind the media rinse reservoir. Ensure the RS232 port is positioned
to the right for a 6-row configuration and toward the back for a
12-row configuration.
2 Connect the nine-pin RS232 cable between the CALIBRATION port
on the 8000 rear panel and the RS232 port on the side of the
autocalibration block.
3 Press H to move the dispensing arm to the home position.
4 Press ESC three times to return to the Ready screen.