lpErrorDoc.txt
Due to hardware limitations, only a certain number of unique data
points can be created for the IQ map. IQ rotation is designed to
alleviate this problem, but incorrect use can result in undesirable
effects. This error indicates that the current IQ rotation is
optimal and cannot be changed.
-223 Too much data; The number of carriers exceeds the maximum allowed.
The maximum number of carriers cannot exceed the maximum that is
defined in the technical specifications document. Reduce the number
of carriers and try again.
-223 Too much data; The number of CDMA carriers exceeds the maximum
allowed.
The maximum number of CDMA carriers cannot exceed the maximum that
is defined in the technical specifications document. Reduce the
number of channels and try again.
-223 Too much data; The number of CDMA channels exceeds the maximum
allowed.
The maximum number of CDMA channels cannot exceed the maximum that
is defined in the technical specifications document. Reduce the
number of carriers and try again.
-223 Too much data; The number of list points exceeds the maximum allowed.
Some tables have a maximum number of entries. Sweep/List can only
have 1601 points, for example. The baseband generator cannot use an
I/Q map that has more than 16 unique values for both I and Q. You
can use global search and replace to replace all occurrences of a
given value (in both I and Q) with a different value.
-223 Too much data; The number of points in the table exceeds the maximum
allowed.
This error occurs when you try to create a table that is too big.
For example, if you enter a list in SCPI that is longer than the
maximum allowed length, or if you attempt to expand a table and the
table is already at its maximum length.
-223 Too much data; User filter has too many symbols.
The baseband generator cannot use a filter that has more than 64
symbols. You may have specified an incorrect over-sample ratio in
the filter table editor.
-223 Too much data; {name}
The specified file was being built up by a SCPI data block transfer
when the memory block allocation failed.
-222 Data out of range
A legal program data element was parsed but could not be
executed because the interpreted value was outside the legal
range defined by the device (see IEEE 488.2 11.5.1.1.5).
-222 Data out of range; A burst fall shape must have at least two points.
Two points are needed to specify a fall shape in the MEM:DATA:SHAPe
Page 16